Remote Control
R&S
®
FSVA/FSV
629Operating Manual 1176.7510.02 ─ 09
CALCulate<n>:DELTamarker<m>:SGRam:XY:MAXimum[:PEAK]
This command positions the specified delta marker on the current peak level of the
spectrogram. The possible position depends on the deltamarker search area.
The command is available for the spectrogram.
For more information, see CALCulate<n>:DELTamarker<m>:SGRam:SARea
on page 628.
Suffix:
<n>
.
1...4
Selects the measurement window.
<m> 1...16
Selects the marker.
Example:
CALC:DELT2:SGR:SAR VIS
CALC:DELT2:SGR:XY:MAX
Activates and positions delta marker 2 on the peak level in the
visible Spectrogram.
Usage: Event
Manual operation: See "Marker Search Type" on page 306
CALCulate<n>:DELTamarker<m>:SGRam:XY:MINimum[:PEAK]
This command positions the specified delta marker on the current minimum level of the
spectrogram. The possible position depends on the deltamarker search area.
The command is available for the spectrogram.
For more information, see CALCulate<n>:DELTamarker<m>:SGRam:SARea
on page 628.
Suffix:
<n>
.
1...4
Selects the measurement window.
<m> 1...16
Selects the marker.
Example:
CALC:DELT3:SGR:SAR VIS
CALC:DELT3:SGR:XY:MIN
Activates and positions delta marker 3 on the minimum level in
the visible Spectrogram.
Usage: Event
Manual operation: See "Marker Search Type" on page 306
Remote Control – Commands