Remote Control
R&S
®
FSVA/FSV
628Operating Manual 1176.7510.02 ─ 09
Suffix:
<n>
.
1...4
Selects the measurement window.
<m> 1...16
Selects the marker.
Parameters:
<Frame> | <Time> <Frame> (if time stamp is off)
Defines the distance of the deltamarker in frames and in relation
to the position of marker 1.
<time> (if time stamp is on)
Defines the distance of the deltamarker in seconds and in rela-
tion to the position of marker 1. The reference is the time stamp
of marker 1.
Example:
CALC:DELT4:SGR:FRAM -20
Sets fourth deltamarker 20 frames below marker 1.
CALC:DELT4:SGR:FRAM 2 s
Sets fourth deltamarker 2 seconds above the position of marker
1.
CALCulate<n>:DELTamarker<m>:SGRam:SARea <SearchArea>
This command defines the deltamarker search area.
The command is available for the spectrogram.
Suffix:
<n>
.
1...4
Selects the measurement window.
<m> 1...16
Selects the marker.
Parameters:
<SearchArea> VISible
Performs a search in the visible frames.
Note that VISible is ignored if Spectrogram is not visible for any
reason (e.g. if the Spectrum Analyzer is in full screen mode or
display update is inactive).
MEMory
Performs a search over all frames in the memory.
*RST: VISible
Example:
CALC:DELT:SGR:SAR MEM
Performs a search over all frames in the memory.
Manual operation: See "Marker Search Area" on page 306
Remote Control – Commands

Table of Contents