EasyManua.ls Logo

Rohde & Schwarz FSVA - Page 674

Rohde & Schwarz FSVA
1106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
R&S
®
FSVA/FSV
674Operating Manual 1176.7510.02 ─ 09
Example:
CALC:MARK2:X 1.7MHz
Positions marker 2 to frequency 1.7 MHz.
Manual operation: See "Marker 1 / Marker 2 / Marker 3 / … Marker 16,/ Marker
Norm/Delta" on page 280
CALCulate<n>:MARKer<m>:X:SLIMits[:STATe] <State>
This command turns marker search limits on and off.
If the power measurement in zero span is active, this command limits the evaluation
range on the trace.
Suffix:
<n>
.
Selects the measurement window.
<m> marker
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:MARK:X:SLIM ON
Switches on search limitation.
Manual operation: See "Search Limits" on page 303
See "Search Lim Off" on page 304
See "Limits (On/Off)" on page 385
CALCulate<n>:MARKer<m>:X:SLIMits:LEFT <Limit>
This command sets the left limit of the marker search range.
If the power measurement in zero span is active, this command limits the evaluation
range to the trace.
Note: The function is only available if the search limit for marker and delta marker is
switched on (see CALCulate<n>:MARKer<m>:X:SLIMits[:STATe]).
Suffix:
<n>
.
Selects the measurement window.
<m> irrelevant
Parameters:
<Limit> The value range depends on the span or sweep time.
The unit is Hz for frequency domain measurements and s for
time domain measurements.
Range: 0 to MAX
*RST: left diagram border
Example:
CALC:MARK:X:SLIM ON
Switches the search limit function on.
CALC:MARK:X:SLIM:LEFT 10MHz
Sets the left limit of the search range to 10 MHz.
Remote Control – Commands

Table of Contents

Related product manuals