Instrument Functions
R&S
®
FSVA/FSV
385Operating Manual 1176.7510.02 ─ 09
Time Domain Power
Activates the power measurement in zero span and opens a submenu to configure the
power measurement. For more details see also Chapter 3.3.5.7, "Measuring the Power
in Zero Span", on page 383.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary[:STATe] on page 727
Peak ← Time Domain Power
Activates the calculation of the peak value from the points of the displayed trace or a
segment thereof.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PPEak[:STATe] on page 721
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PPEak:RESult? on page 721
RMS ← Time Domain Power
Activates the calculation of the RMS value from the points of the displayed trace or a
segment thereof.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:RMS[:STATe] on page 724
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:RMS:RESult? on page 723
Mean ← Time Domain Power
Activates the calculation of the mean value from the points of the displayed trace or a
segment thereof. The linear mean value of the equivalent voltages is calculated.
This can be used for instance to measure the mean power during a GSM burst.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:MEAN[:STATe] on page 719
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:MEAN:RESult? on page 718
Std Dev ← Time Domain Power
Activates the calculation of the standard deviation of measurement points from the
mean value and displays them as measured value. The measurement of the mean
power is automatically switched on at the same time.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:SDEViation[:STATe]
on page 726
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:SDEViation:RESult?
on page 726
Limits (On/Off) ← Time Domain Power
Switches the limitation of the evaluation range on or off. Default setting is off.
If switched off, the evaluation range is not limited. If switched on, the evaluation range
is defined by the left and right limit. If only one limit is set, it corresponds to the left limit
and the right limit is defined by the stop frequency. If the second limit is also set, it
defines the right limit.
Remote command:
CALCulate<n>:MARKer<m>:X:SLIMits[:STATe] on page 674
Measurement Functions