CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:SDEViation:RESult?
This command queries the standard deviation of the time domain power.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps.
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Return values:
<StandardDeviation> Standard deviation of the signal during the measurement time.
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK:FUNC:SUMM:SDEV ON
Switches on the function.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:FUNC:SUMM:SDEV:RES?
Outputs the result.
Usage: Query only
Manual operation: See "Std Dev" on page 385
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:SDEViation[:STATe] <State>
This command turns the measurement of the standard deviation of the time domain
power on and off.
Note: The measurement is performed on the trace marker 1 is positioned. In order to
evaluate another trace, marker 1 must be positioned on another trace with
CALCulate<n>:MARKer<m>:TRACe.
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:MARK:FUNC:SUMM:SDEV ON
Switches on the measurement of the standard deviation.
Manual operation: See "Std Dev" on page 385