Remote Control
R&S
®
FSVA/FSV
724Operating Manual 1176.7510.02 ─ 09
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps.
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Return values:
<RMSPower> RMS power of the signal during the measurement time.
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK:FUNC:SUMM:RMS ON
Switches on the function.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:FUNC:SUMM:RMS:RES?
Outputs the result.
Usage: Query only
Manual operation: See "RMS" on page 385
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:RMS[:STATe] <State>
This command turns the measurement of the RMS time domain power on and off.
Note: The measurement is performed on the trace marker 1 is positioned. In order to
evaluate another trace, marker 1 must be positioned on another trace with
CALCulate<n>:MARKer<m>:TRACe.
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:MARK:FUNC:SUM:RMS ON
Switches on the function.
Manual operation: See "RMS" on page 385
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:SDEViation:AVERage:RESult?
This command queries the average standard deviation of the time domain power. The
query is only possible if averaging has been activated previously using
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:AVERage.
Remote Control – Commands