Remote Control
R&S
®
FSVA/FSV
716Operating Manual 1176.7510.02 ─ 09
Example:
DISP:WIND:TRAC:Y:RLEV -10dBm
Sets the reference level to 10 dB
INP:ATT 30 dB
Sets the input attenuation to 30 dB
FREQ:CENT 935.2MHz;SPAN 0 Hz
Sets the receive frequency to 935.2 MHz and the span to 0 Hz
BAND:RES 1 MHz;VID 3 MHz
Sets the resolution bandwidth to 1 MHz and the video bandwidth
to 3 MHz
DET RMS
Sets the RMS detector
TRIG:SOUR VID;LEV:VID 50 PCT
Selects the trigger source VIDeo and sets the level of the video
trigger source to 50 PCT
SWE:TIME 50 ms
Sets the sweep time to 50 ms
INIT;*WAI
Starts the measurement with synchronization
CALC:MARK:FUNC:MSUM? 50US,450US,576.9US,8
Queries 8 bursts with an offset of 50 µs, a test time of 450 µs
and a period of 576.9 µs
Usage:
Query only
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:AOFF
This command turns all time domain power measurements off.
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Example:
CALC:MARK:FUNC:SUMM:AOFF
Switches off the functions for power measurement in zero span.
Usage: Event
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:AVERage <State>
This command turns averaging for the active power measurement in zero span on and
off.
Averaging is reset by switching it off and on again.
The number of results required for the calculation of average is defined with [SENSe:
]AVERage<n>:COUNt.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps.
Remote Control – Commands