Remote Control
R&S
®
FSVA/FSV
780Operating Manual 1176.7510.02 ─ 09
[SENSe:]AVERage<n>[:STATe<Trace>] <State>
This command turns averaging for a particular trace in a particular window on and off.
Suffix:
<n>
.
Selects the measurement window.
<Trace> 1...6
Selects the trace.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
AVER OFF
Switches off the average calculation for trace 1.
AVER:STAT3 ON
Switches on the average calculation for trace 3.
[SENSe:]AVERage<n>:COUNt
<NoMeasurements>
This command defines the number of measurements which contribute to the average
value.
Note that continuous averaging is performed after the indicated number has been
reached in continuous sweep mode.
In single sweep mode, the sweep is stopped as soon as the indicated number of mea-
surements (sweeps) is reached. Synchronization to the end of the indicated number of
measurements is only possible in single sweep mode.
This command has the same effect as the [SENSe<source>:]SWEep:COUNt com-
mand. In both cases, the number of measurements is defined whether the average cal-
culation is active or not.
The number of measurements applies to all traces in the window.
Suffix:
<n>
.
Selects the measurement window.
Parameters:
<NoMeasurements> 0 to 32767
*RST: 0
Example:
SWE:CONT OFF
Switching to single sweep mode.
AVER:COUN 16
Sets the number of measurements to 16.
AVER:STAT ON
Switches on the calculation of average.
INIT;*WAI
Starts the measurement and waits for the end of the 16 sweeps.