Remote Control
R&S
®
FSVA/FSV
715Operating Manual 1176.7510.02 ─ 09
CALCulate<n>:MARKer<m>:FUNCtion:MSUMmary?
The commands of this subsystem are used to determine the power of a sequence of
signal pulses that have the same interval in the window specified by the suffix <n>, as
for example the slots of a GSM signal typically do. The number of pulses to be mea-
sured as well as the measurement time and the period can be set. To define the posi-
tion of the first pulse in the trace, a suitable offset can be entered.
The evaluation is performed on the measurement data of a previously recorded trace.
The data recorded during the set measurement time is combined to a measured value
for each pulse according to the detector specified and the indicated number of results
is output as a list.
Trace 1 is always used by the function.
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Parameters:
<time offset of first pulse>, <measurement time>, <period>, < #
of pulses to measure>