624Operating Manual 1176.7510.02 ─ 09
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Example:
CALC:DELT:MIN:LEFT
Sets delta marker 1 to the next higher minimum to the left of the
current value.
Manual operation: See "Next Mode" on page 304
CALCulate<n>:DELTamarker<m>:MINimum:RIGHt
This command positions the delta marker to the next higher trace minimum on the right
of the current value (i.e. ascending X values). The corresponding delta marker is acti-
vated first, if necessary.
If no next higher minimum value is found on the trace (level spacing to adjacent values
< peak excursion), an execution error (error code: -200) is produced.
In "Spectrogram" mode, this command positions the specified delta marker on the next
minimum level of the x-axis. The search includes only the data to the right of the cur-
rent marker position in the current frame. It does not change the position of the delta
marker on the y-axis.
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Example:
CALC:DELT:MIN:RIGH
Sets delta marker 1 to the next higher minimum value to the
right of the current value.
Manual operation: See "Next Mode" on page 304
CALCulate<n>:DELTamarker<m>:MINimum:NEXT
This command positions the delta marker to the next higher trace minimum. The corre-
sponding delta marker is activated first, if necessary.
If no next higher minimum value is found on the trace (level spacing to adjacent values
< peak excursion), an execution error (error code: -200) is produced.
In "Spectrogram" mode, this command positions the specified delta marker on the next
minimum level of the x-axis. The search includes the data in both directions of the cur-
rent marker position in the current frame. It does not change the position of the delta
marker on the y-axis.
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Example:
CALC:DELT2:MIN:NEXT
Sets delta marker 2 to the next higher minimum value.