Loading...
Remote Control
R&S
®
FSVA/FSV
965Operating Manual 1176.7510.02 ─ 09
SYSTem:DISPlay:LANGuage <Language>
Changes the language of the graphical user interface (softkeys, dialog boxes, input
settings etc.) to the selected language, if any other than the default are installed.
Parameters:
<Language> string containing the ISO language code
*RST: en
Example:
SYST:DISP:LANG 'ru'
Switches to a Russian interface (if installed).
SYST:DISP:LANG 'en'
Switches back to the default English interface.
Manual operation: See "Language" on page 185
SYSTem:DISPlay:UPDate <State>
In remote control mode, this command switches on or off the instrument display. If
switched on, only the diagrams, traces and display fields are displayed and updated.
The best performance is obtained if the display output is switched off during remote
control.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
SYST:DISP:UPD ON
Manual operation: See "Display Update (On/Off)" on page 184
SYSTem:ERRor:CLEar:ALL
This command deletes all entries in the table SYSTEM MESSAGES.
Example:
SYST:ERR:CLE:ALL
Usage: Event
Manual operation: See "Clear All Messages" on page 191
SYSTem:ERRor:LIST?
This command reads all system messages and returns a list of comma separated
strings. Each string corresponds to an entry in the table SYSTEM MESSAGES.
If the error list is empty, an empty string " is returned.
This command is a query and therefore has no *RST value.
Example:
SYST:ERR:LIST?
Usage: Query only
Remote Control – Commands