Instrument Functions
R&S
®
FSVA/FSV
184Operating Manual 1176.7510.02 ─ 09
Display Update (On/Off) ← GPIB ← General Setup
Defines whether the instrument display is switched off when changing from manual
operation to remote control. In remote control mode, this softkey is displayed in the
local menu.
Remote command:
SYSTem:DISPlay:UPDate on page 965
GPIB Terminator LFEOI/EOI ← GPIB ← General Setup
Changes the GPIB receive terminator.
According to the standard, the terminator in ASCII is <LF> and/or <EOI>. For binary
data transfers (e.g. trace data) from the control computer to the instrument, the binary
code used for <LF> might be included in the binary data block, and therefore should
not be interpreted as a terminator in this particular case. This can be avoided by
changing the receive terminator to EOI.
Remote command:
SYSTem:COMMunicate:GPIB[:SELF]:RTERminator on page 962
*IDN Format Leg./New ← GPIB ← General Setup
Defines the response format to the *IDN? remote command (see *IDN?
on page 608). This function is intended for re-use of existing control programs
together with the R&S FSVA/FSV.
"Leg"
Legacy format, compatible to the R&S FSP/FSU/FSQ family
e.g. Rohde&Schwarz,FSV-7,100005/007,1.61
"New"
R&S FSVA/FSV format
e.g. Rohde&Schwarz,FSV-7,1307.9002K07/100005,1.61
Remote command:
SYSTem:FORMat:IDENt on page 966
I/O Logging (On/Off) ← GPIB ← General Setup
Activates or deactivates the SCPI log function. All remote control commands received
by the R&S FSVA/FSV are recorded in the following log file:
C:\R_S\Instr\scpilogging\ScpiLog.txt
Logging the commands may be extremely useful for debug purposes, e.g. in order to
find misspelled keywords in control programs.
Remote command:
SYSTem:CLOGging on page 961
Time+Date ← General Setup
Opens an edit dialog box to enter time and date for the internal real time clock. For
details refer to the Quick Start Guide, chapter 2 "Preparing for Use".
Remote command:
SYSTem:TIME on page 972