Instrument Functions
R&S
®
FSVA/FSV
183Operating Manual 1176.7510.02 ─ 09
Remote command:
SYSTem:HPCoupling on page 967
REV String Factory ← Compatibility Mode ← GPIB ← General Setup
Selects the default response to the REV? query for the revision number (HP emulation
only, see Chapter 4.2.5, "GPIB Commands of HP Models 856xE, 8566A/B, 8568A/B
and 8594E", on page 974).
Remote command:
SYSTem:REVision:FACTory on page 971
REV String User ← Compatibility Mode ← GPIB ← General Setup
Opens an edit dialog box to enter a user-defined revision number in response to the
REV? query (HP emulation only, see Chapter 4.2.5, "GPIB Commands of HP Models
856xE, 8566A/B, 8568A/B and 8594E", on page 974). Max. 36 characters are allowed.
Remote command:
SYSTem:REVision[:STRing] on page 972
Tree Walking ← Compatibility Mode ← GPIB ← General Setup
Defines how remote control programs are evaluated, in particular multiple commands
in one command line.
Note that this setting is retained after a PRESET or recall operation.
See the SCPI command for an example.
"Normal"
(Initial default:) SCPI-conform behaviour
The first command is always referenced to the root node. Subsequent
commands, however, are referenced to the same tree level as the
previous command in a message unit.
"Enhanced"
If subsequent commands are not found on the same tree level as the
previous command, the SCPI parser searches "backwards up the
tree" of defined SCPI commands. If the command is found at a higher
node, that command is used. Only if the command is not found any-
where, an error message is returned.
Remote command:
SYSTem:TWALking on page 973
Wideband ← GPIB ← General Setup
This setting defines which option is returned when the *OPT? query is executed,
depending on the state of the wideband option.
It is only available for PSA89600 emulation.
"Wideband
Off"
No wideband is used.
The option is indicated as "B7J".
"Wideband 40
MHz"
The 40 MHz wideband is used.
The option is indicated as "B7J, 140".
"Wideband 80
MHz"
The 80 MHz wideband is used.
The option is indicated as "B7J, 122".
Remote command:
SYSTem:PSA:WIDeband on page 971
General Settings, Printout and Instrument Settings