Instrument Functions
R&S
®
FSVA/FSV
182Operating Manual 1176.7510.02 ─ 09
Language Comment
8568B Command sets A and B are available. Command sets A and B differ in the rules
regarding the command structure.
8568B_DC Uses DC input coupling by default if supported by the instrument
8591E Compatible to 8594E
8594E Command sets A and B are available. Command sets A and B differ in the rules
regarding the command structure.
For details on the GPIB languages, see Chapter 4.2.5, "GPIB Commands of HP Mod-
els 856xE, 8566A/B, 8568A/B and 8594E", on page 974.
Remote command:
SYSTem:LANGuage on page 968
IF Gain (Norm/Puls) ← Compatibility Mode ← GPIB ← General Setup
Configures the internal IF gain settings in HP emulation mode due to the application
needs. This setting is only taken into account for resolution bandwidth < 300 kHz.
NORM Optimized for high dynamic range, overload limit is close to reference level.
PULS Optimized for pulsed signals, overload limit up to 10 dB above reference level.
This softkey is only available if a HP language is selected via the "GPIB Language"
softkey (see "GPIB Language" on page 181).
Remote command:
SYSTem:IFGain:MODE on page 968
Sweep Repeat (On/Off) ← Compatibility Mode ← GPIB ← General Setup
Controls a repeated sweep of the E1 and MKPK HI/HL HP model commands (for
details on the commands refer to "GPIB Language" on page 181). If the repeated
sweep is OFF, the marker is set without sweeping before.
Note: In single sweep mode, switch off this softkey before you set the marker via the
E1 and MKPK HI/HL commands in order to avoid sweeping again.
This softkey is only available if a HP language is selected via the "GPIB Language"
softkey (see "GPIB Language" on page 181).
Remote command:
SYSTem:RSW on page 972
Coupling (FSx/HP) ← Compatibility Mode ← GPIB ← General Setup
Controls the default coupling ratios in the HP emulation mode for:
●
span and resolution bandwidth (Span/RBW) and
●
resolution bandwidth and video bandwidth (RBW/VBW)
For FSP(=FSV), the standard parameter coupling of the instrument is used. As a
result, in most cases a shorter sweeptime is used than in case of HP.
This softkey is only available if a HP language is selected via the "GPIB Language"
softkey (see "GPIB Language" on page 181).