EasyManua.ls Logo

Rohde & Schwarz FSVA - Page 972

Rohde & Schwarz FSVA
1106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
R&S
®
FSVA/FSV
972Operating Manual 1176.7510.02 ─ 09
SYSTem:REVision[:STRing] <Name>
Sets the response to the REV? queryto the defined string (HP emulation only, see
Chapter 4.2.5, "GPIB Commands of HP Models 856xE, 8566A/B, 8568A/B and
8594E", on page 974).
Parameters:
<Name>
Example: Define the system language:
SYST:LANG '8563E'
Query the revision:
REV?
Response:
920528
Set the response to 'NewRevision':
SYST:REV:STR 'NewRevision'
Query the response:
SYST:REV:STR?
Response:
NewRevision
Manual operation:
See "REV String User" on page 183
SYSTem:RSW <State>
Controls a repeated sweep of the E1 and MKPK HI/HL HP model commands (for
details on the commands refer to Chapter 4.2.5, "GPIB Commands of HP Models
856xE, 8566A/B, 8568A/B and 8594E", on page 974). If the repeated sweep is OFF,
the marker is set without sweeping before.
This command is only available if a HP language is selected using SYSTem:LANGuage
.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
SYSTem:RSW ON
Usage: SCPI confirmed
Manual operation: See "Sweep Repeat (On/Off)" on page 182
SYSTem:SHUTdown
This command shuts down the instrument.
Usage:
Event
SYSTem:TIME <Time>
This command sets the internal clock. The sequence of entry is hour, minute, second.