Remote Control
R&S
®
FSVA/FSV
968Operating Manual 1176.7510.02 ─ 09
Manual operation: See "ID String User" on page 180
SYSTem:IFGain:MODE <Mode>
Configures the internal IF gain settings in HP emulation mode due to the application
needs. This setting is only taken into account for resolution bandwidth < 300 kHz and is
only available if a HP language is selected using SYSTem:LANGuage.
Parameters:
<Mode> NORM | PULS
NORM
Optimized for high dynamic range, overload limit is close to ref-
erence level.
PULS
Optimized for pulsed signals, overload limit up to 10 dB above
reference level.
*RST: NORM
Example:
SYST:IFG:MODE PULS
Usage: SCPI confirmed
Manual operation: See "IF Gain (Norm/Puls)" on page 182
SYSTem:KLOCk <State>
This command activates the local lockout (remote control) or returns to the local mode.
Parameters:
<State> ON
LLO (local lockout)
OFF
GTL (go to local)
*RST: OFF
Example:
SYST:KLOK ON
Activates LLO (remote control)
SYSTem:LANGuage <Language>
This command defines the system language. Default is SCPI which defines the stan-
dard remote control behavior of the instrument. The other parameters are to simulate
the corresponding HP analyzer. The parameter "PSA89600" sets the device in a state
in which it can be used for IQ capturing from the 89600 vector signal analyzer soft-
ware.
Remote Control – Commands