SYSTem:HPCoupling <CouplingType>
Controls the default coupling ratios in the HP emulation mode for:
span and resolution bandwidth (Span/RBW) and
resolution bandwidth and video bandwidth (RBW/VBW)
For FSP (=FSV), the standard parameter coupling of the instrument is used. As a
result, in most cases a shorter sweeptime is used than in case of HP.
This command is only available if a HP language is selected using SYSTem:LANGuage
.
Parameters:
<CouplingType> HP | FSP
*RST: FSP
Example:
SYSTem:HPC HP
Usage: SCPI confirmed
Manual operation: See "Coupling (FSx/HP)" on page 182
SYSTem:IDENtify:FACTory
Resets the response to the *IDN? query to the default value, e.g. after a user string
was defined using the ID String User softkey.
Example:
SYST:IDEN:FACT
*IDN?
Response:
Rohde&Schwarz,R&S&#x00a0;FSVA/FSV-7,
1307.9002K07/101768,1.05
Usage: Event
SCPI confirmed
Manual operation: See "ID String Factory" on page 180
SYSTem:IDENtify[:STRing] <String>
Sets the response to the *IDN? query to the defined string, e.g. after the default value
was reset using the ID String Factory softkey.
Parameters:
<String>
Example: Set the response to 'MyDevice':
SYST:IDEN:STR 'MyDevice'
Query the response:
SYST:IDEN:STR?
Response:
MyDevice
Usage: SCPI confirmed