Remote Control
R&S
®
FSVA/FSV
966Operating Manual 1176.7510.02 ─ 09
Manual operation: See "System Messages" on page 191
SYSTem:ERRor[:NEXT]?
This command queries the earliest error queue entry and deletes it.
Positive error numbers indicate device-specific errors, negative error numbers are error
messages defined by SCPI. If the error queue is empty, the error number 0, "No error",
is returned.
This command is a query and therefore has no *RST value.
Example:
STAT:ERR?
Usage: Query only
SCPI confirmed
SYSTem:FIRMware:UPDate
This command starts a firmware update using the *.msi files in the selected directory.
The default path is D:\FW_UPDATE. The path is changed via the MMEMory:COMMent
command. To store the update files the MMEMory:DATA command is used.
Only user accounts with administrator rights can perform a firmware update.
Example:
SYST:FIRM:UPD 'D:\FW_UPDATE'
Starts the firmware update from directory "D:\FW_UPDATE".
Usage: Event
Manual operation: See "Firmware Update" on page 191
SYSTem:FORMat:IDENt
This command defines the response format to the *IDN? on page 608 command. This
function is intended for re-use of existing control programs together with the
R&S FSVA/FSV.
Parameters:
FSL | LEGacy | NEW
LEGacy
Format is compatible to the R&S FSP/FSU/FSQ family
NEW
R&S FSVA/FSV format
FSL
Format is compatible to the R&S FSL family
*RST: not reset!
Example:
SYST:FORM:IDEN LEG
Adapts the return value of *IDN? to the R&S FSP/FSU/FSQ fam-
ily.
Manual operation: See "*IDN Format Leg./New" on page 184
Remote Control – Commands