EasyManua.ls Logo

Rohde & Schwarz FSVA - Page 966

Rohde & Schwarz FSVA
1106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
R&S
®
FSVA/FSV
966Operating Manual 1176.7510.02 ─ 09
Manual operation: See "System Messages" on page 191
SYSTem:ERRor[:NEXT]?
This command queries the earliest error queue entry and deletes it.
Positive error numbers indicate device-specific errors, negative error numbers are error
messages defined by SCPI. If the error queue is empty, the error number 0, "No error",
is returned.
This command is a query and therefore has no *RST value.
Example:
STAT:ERR?
Usage: Query only
SCPI confirmed
SYSTem:FIRMware:UPDate
This command starts a firmware update using the *.msi files in the selected directory.
The default path is D:\FW_UPDATE. The path is changed via the MMEMory:COMMent
command. To store the update files the MMEMory:DATA command is used.
Only user accounts with administrator rights can perform a firmware update.
Example:
SYST:FIRM:UPD 'D:\FW_UPDATE'
Starts the firmware update from directory "D:\FW_UPDATE".
Usage: Event
Manual operation: See "Firmware Update" on page 191
SYSTem:FORMat:IDENt
This command defines the response format to the *IDN? on page 608 command. This
function is intended for re-use of existing control programs together with the
R&S FSVA/FSV.
Parameters:
FSL | LEGacy | NEW
LEGacy
Format is compatible to the R&S FSP/FSU/FSQ family
NEW
R&S FSVA/FSV format
FSL
Format is compatible to the R&S FSL family
*RST: not reset!
Example:
SYST:FORM:IDEN LEG
Adapts the return value of *IDN? to the R&S FSP/FSU/FSQ fam-
ily.
Manual operation: See "*IDN Format Leg./New" on page 184