Instrument Functions
R&S
®
FSVA/FSV
180Operating Manual 1176.7510.02 ─ 09
Remote command:
SYSTem:LXI:MDEScription on page 969
LAN Reset ← LXI ← General Setup
Resets the LAN configuration to a state required by the LXI standard. For example, the
TCP/IP mode is set to DHCP and Dynamic DNS and ICMP Ping are enabled. In addi-
tion, the R&S FSVA/FSV sets the password and the instrument description to their ini-
tial states (see "Password" on page 179 and "Description" on page 179 softkeys).
Only user accounts with administrator rights can reset the LAN configuration.
Remote command:
SYSTem:LXI:LANReset on page 969
GPIB ← General Setup
Opens a submenu to set the parameters of the remote control interface.
GPIB Address ← GPIB ← General Setup
Opens an edit dialog box to enter the GPIB address. Values from 0 to 30 are allowed.
The default address is 20.
Remote command:
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess on page 961
ID String Factory ← GPIB ← General Setup
Selects the default response to the *IDN? query.
Remote command:
SYSTem:IDENtify:FACTory on page 967
ID String User ← GPIB ← General Setup
Opens an edit dialog box to enter a user-defined response to the *IDN? query. Max. 36
characters are allowed.
Remote command:
SYSTem:IDENtify[:STRing] on page 967
Compatibility Mode ← GPIB ← General Setup
Sets the R&S FSVA/FSV in a state compatible to previous R&S devices, enabling the
usage of existing external control applications. In particular, the number of measure-
ment points and available bandwidths are adjusted to those of other devices.
Furthermore, some special GPIB settings are available in order to emulate HP models
(see Chapter 4.2.5, "GPIB Commands of HP Models 856xE, 8566A/B, 8568A/B and
8594E", on page 974):
●
"GPIB Language" on page 181
●
"IF Gain (Norm/Puls)" on page 182
●
"Sweep Repeat (On/Off)" on page 182
●
"Coupling (FSx/HP)" on page 182
"Default"
Standard R&S FSVA/FSV operation, see "Mode Default"
on page 181
"R&S FSP"
Compatible to R&S FSP, see "Mode R&S FSP" on page 181
General Settings, Printout and Instrument Settings