Remote Control
R&S
®
FSVA/FSV
961Operating Manual 1176.7510.02 ─ 09
SYSTem:PSA:WIDeband................................................................................................971
SYSTem:REBoot........................................................................................................... 971
SYSTem:REVision:FACTory........................................................................................... 971
SYSTem:REVision[:STRing]............................................................................................972
SYSTem:RSW...............................................................................................................972
SYSTem:SHUTdown......................................................................................................972
SYSTem:TIME...............................................................................................................972
SYSTem:TWALking....................................................................................................... 973
SYSTem:VERSion?....................................................................................................... 973
SYSTem:APPLication:SRECovery[:STATe] <State>
Controls instrument behavior when switching between measurement applications, e.g.
from "Spectrum" to "Analog Demod" and back.
If activated, the current instrument settings are stored when you switch to a different
application. When you switch back to the previous application, the corresponding
instrument settings are restored. Thus, the settings of the individual applications are
independant of each other.
If deactivated (default), only a few parameters of the current instrument setting are
passed between applications (e.g. center frequency, level settings).
Note that although the default value is "OFF", this parameter is not reset using the
PRESET key. Thus it must be deactivated manually, if necessary.
Parameters:
<State> ON | OFF
*RST: OFF (not restored using the "PRESET" function)
Example:
SYST:APPL:SREC ON
Manual operation: See "Application Setup Recovery" on page 192
SYSTem:CLOGging <State>
This command turns logging of remote commands on and off.
Parameters:
<State> ON
Writes all remote commands that have been sent to a file.
The destination is C:
\R_S\instr\ScpiLogging\ScpiLog.txt.
OFF
*RST: OFF
Manual operation: See "I/O Logging (On/Off)" on page 184
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess <Address>
This command changes the GPIB address of the unit.
Remote Control – Commands