Remote Control
R&S
®
FSVA/FSV
962Operating Manual 1176.7510.02 ─ 09
Parameters:
<Address> 0 to 30
*RST: (no influence on this parameter, factory default 20)
Example:
SYST:COMM:GPIB:ADDR 18
Usage: SCPI confirmed
Manual operation: See "GPIB Address" on page 180
SYSTem:COMMunicate:GPIB[:SELF]:RTERminator <Terminator>
This command changes the GPIB receive terminator.
According to the standard the terminator in ASCII is <LF> and/or <EOI>. For binary
data transfers (e.g. trace data) from the control computer to the instrument, the binary
code (0AH) used for <LF> might be included in the binary data block, and therefore
should not be interpreted as a terminator in this particular case. This can be avoided by
changing the receive terminator to EOI.
Output of binary data from the instrument to the control computer does not require
such a terminator change.
Parameters:
<Terminator> LFEOI | EOI
*RST: (no influence on this parameter, factory default
LFEOI)
Example:
SYST:COMM:GPIB:RTER EOI
Manual operation: See "GPIB Terminator LFEOI/EOI" on page 184
SYSTem:COMMunicate:PRINter:ENUMerate:FIRSt?
This command queries the name of the first printer (in the list of printers) available
under Windows operating system.
The names of other installed printers can be queried with the SYSTem:COMMunicate:
PRINter:ENUMerate[:NEXT]? command.
If no printer is configured an empty string is output.
Parameters:
*RST: NONE
Example:
SYST:COMM:PRIN:ENUM:FIRS?
Usage: Query only
Manual operation: See "Device Setup" on page 212