Remote Control
R&S
®
FSVA/FSV
767Operating Manual 1176.7510.02 ─ 09
Mode Text parameter
(INST:SEL)
Numeric
parameter
(INST:NSEL)
Option number
GSM GSM
(Query: MGSM.)
5 R&S FSV-K10
Noise Figure Measurements NOISe 19 R&S FSV-K30
Phase Noise mode PNOise 20 R&S FSV-K40
VSA DDEM 2 R&S FSV-K70
3G FDD BTS mode BWCD 8 R&S FSV-K72
3G FDD UE mode MWCD 9 R&S FSV-K73
TD-SCDMA BTS mode BTDS 17 R&S FSV-K76
TD-SCDMA UE mode MTDS 18 R&S FSV-K77
CDMA2000 BS Analysis BCK2 10 R&S FSV-K82
CDMA2000 MS MC2K 11
1xEV-DO BS Analysis BDO 14 R&S FSV-K84
1xEV-DO MS MDO 15
WLAN TX WLAN 16 R&S FSV-K91/91n
WiMax WiMAX 23 R&S FSV-K93
LTE (uplink and downlink) LTE 100 R&S FSV-K101/K102/K103/K104/
K105
INSTrument:CREate[:NEW] <ChannelType>, <ChannelName>
This command adds an additional spectrum display. You can add up to three additional
spectrum displays.
Also see
●
INSTrument[:SELect] on page 766
●
INSTrument:DELete on page 768
●
Remote program example: Chapter 4.3.16, "Usage of Four Spectrum Instances",
on page 1049
Parameters:
<ChannelType> SANalyzer
The channel type is always SANalyzer to indicate that you add
a new spectrum display.
<ChannelName> Sets the name of the additional spectrum display.
Spectrum 2 | Spectrum 3 | Spectrum 4
The names of the spectrum displays you add have to be
Spectrum 2 for the second spectrum display, Spectrum 3 for
the third and Spectrum 4 for the fourth.
Remote Control – Commands