Example:
SENSe:LIST:POWer
935.2 MHz,0dBm,10dB,OFF,NORM,1 MHz,3 MHz,440us,
0,
935.4 MHz,0dBm,10dB,10dB,CFIL,30 kHz,100 kHz,
440us,0,
935.6 MHz,0dBm,10dB,20dB,CFIL,30 kHz,100 kHz,
440us,0
Performs a measurement sequence with the settings described
in Table 4-15.
SENSe:LIST:POWer?
935.2 MHz,0dBm,10dB,OFF,NORM,1 MHz,3 MHz,440us,
0,
935.4 MHz,0dBm,10dB,10dB,CFIL,30 kHz,100 kHz,
440us,0,
935.6 MHz,0dBm,10dB,20dB,CFIL,30 kHz,100 kHz,
440us,0
Performs the same measurement and returns the result list
immediately after the last frequency point.
Table 4-15: Settings for power list example
Step Freq.
[MHz]
Ref
Level
[dBm]
RF Att
[dB]
el Att
[dB]
Filter
type
RBW VBW Meas
Time
[us]
TRG
Level
(reserve
d)
1 935.2 0 10 OFF Normal 1 MHz 3 MHz 440 0
2 935.4 0 10 10 Channel 30 kHz 100 kHz 440 0
3 935.6 0 10 20 Channel 30 kHz 100 kHz 440 0
[SENSe:]LIST:POWer:SET
<PeakPower>,<RMSPower>,<AVGPower>,<TriggerSource>,<TriggerSlope>,<T
riggerOffset>,<GateLength>
This command defines global List Evaluation parameters.
These parameters are valid for every frequency you want to measure.
The state of the first three parameters (<PeakPower>, <RMSPower> and <AVG-
Power>) define the number of results for each frequency in the list.
Note that you have to set the trigger level after sending this command.
Parameters:
<PeakPower> ON | OFF
Turns peak power evaluation on and off.
*RST: ON
<RMSPower> ON | OFF
Turns RMS power evaluation on and off.
*RST: OFF
Remote Control – Commands