Remote Control
R&S
®
FSVA/FSV
812Operating Manual 1176.7510.02 ─ 09
<RBW> Defines the resolution bandwidth for a list entry.
For the range of values refer to "Res BW Manual" on page 238.
Refer to Chapter 3.2.6.3, "Selecting the Appropriate Filter Type",
on page 243 for possible combinations of filter type and filter
bandwidth for the <filter type> = CFILter and <filter type> =
RRC.
<VBW> Defines the video bandwidth for a list entry.
The value is ignored for <filter type> = "CFILter" or "RRC"
Range: 1 Hz to 10 MHz in 1, 3, 10 steps
<MeasTime> Defines the measurement time for a list entry.
For details refer to "Sweeptime Manual" on page 239.
Range: 1 µs to 16000 s
<TriggerLevel> The trigger level must be 0.
Return values:
<PowerLevel> Power level for each frequency included in the measurement.
The command returns up to 3 power levels for each frequency,
depending on the number of evaluation modes you have turned
on with [SENSe:]LIST:POWer:SET on page 813.
The result is a list of floating point values separated by commas.
The unit depends on CALCulate<n>:UNIT:POWer.
Example:
SENSe:LIST:POWer?
935.2 MHz,0 dBm,10 dB,OFF,NORM,1 MHz,3 MHz,
440us,0,
935.4 MHz,0dBm,10dB,10dB,NORM,30 kHz,100 kHz,
440us,0,
935.6 MHz,0dBm,10dB,20dB,NORM,30 kHz,100 kHz,
440us,0
Result list:
-28.3,-30.6,-38.1
Example:
SENSe:LIST:POWer:SET ON,ON,ON,IMM,POS,0,0
Activates the power measurement of the peak, RMS and aver-
age values.
SENSe:LIST:POWer?
935.2 MHz,0dBm,10dB,OFF,NORM,1 MHz,3 MHz,440us,
0,
935.4 MHz,0dBm,10dB,10dB,NORM,30 kHz,100 kHz,
440us,0,
935.6 MHz,0dBm,10dB,20dB,NORM,30 kHz,100 kHz,
440us,0
Result list (3 results per frequency point):
-28.3, -29.6, 1.5, -30.6, -31.9, 0.9, -38.1, -40.0, 2.3
Remote Control – Commands