Remote Control
R&S
®
FSVA/FSV
855Operating Manual 1176.7510.02 ─ 09
Parameters:
<Source> EXTernal | IFPower | VIDeo | RFPower | PSEN
*RST: IFPower
Example:
SWE:EGAT:SOUR IFP
Switches the gate source to IF power.
Manual operation: See "Trg/Gate Source" on page 268
See "External" on page 269
See "Video" on page 269
See "RF Power" on page 269
See "IF Power/BB Power" on page 270
See "Power Sensor" on page 270
See "Digital IQ" on page 271
See "Gated Trigger" on page 274
See "Gated Trigger (On/Off)" on page 394
[SENSe:]SWEep:EGATe:TRACe<k>:COMMent <Comment>
Defines a comment for one of the traces for gated triggering.
Suffix:
<k>
.
1...6
trace
Parameters:
<Comment>
Example:
SWE:EGAT:TRAC1:COMM "SlotA"
Manual operation: See "Gate Ranges" on page 394
[SENSe:]SWEep:EGATe:TRACe<k>[:STATe<range>] <State>
This command activates or deactivates tracing for a specific range using gated trigger-
ing.
Suffix:
<k>
.
1...6
trace
<range> 1...3
range
Parameters:
<State> ON | OFF
*RST: OFF
Example:
SWE:EGAT:TRAC1:STAT1 ON
Activates tracing for range 1 of trace 1.
Manual operation: See "Gate Ranges" on page 394
Remote Control – Commands