Remote Control
R&S
®
FSVA/FSV
854Operating Manual 1176.7510.02 ─ 09
[SENSe:]SWEep:EGATe:HOLDoff <DelayTime>
This command defines the delay time between the external gate signal and the contin-
uation of the sweep.
Note: Using gate mode "level" (see [SENSe:]SWEep:EGATe:TYPE on page 857)
and an IFP trigger (see TRIGger<n>[:SEQuence]:SOURce on page 909), the
holdoff time for the IFP trigger is ignored for frequency sweep, FFT sweep, zero span
and IQ mode measurements.
Parameters:
<DelayTime> 0 s to 30 s
*RST: 0s
Example:
SWE:EGAT:HOLD 100us
Manual operation: See "Gate Delay" on page 274
[SENSe:]SWEep:EGATe:LENGth
<TimeInterval>
This command defines a gate length.
Parameters:
<TimeInterval> 125 ns to 30 s
*RST: 400μs
Example:
SWE:EGAT:LENG 10ms
Manual operation: See "Gate Length (Gate Mode Edge)" on page 275
[SENSe:]SWEep:EGATe:POLarity <Polarity>
This command determines the polarity of the external gate signal. The setting applies
both to the edge of an edge-triggered signal and the level of a level-triggered signal.
Parameters:
<Polarity> POSitive | NEGative
*RST: POSitive
Example:
SWE:EGAT:POL POS
Manual operation: See "Trg/Gate Polarity" on page 272
See "Trigger Polarity" on page 462
[SENSe:]SWEep:EGATe:SOURce <Source>
This command selects the signal source for gated measurements.
If an IF power signal is used, the gate is opened as soon as a signal at > -20 dBm is
detected within the IF path bandwidth (10 MHz).
For details see the "Trg/Gate Source" on page 268 softkey.
Remote Control – Commands