Remote Control
R&S
®
FSVA/FSV
909Operating Manual 1176.7510.02 ─ 09
TRIGger<n>[:SEQuence]:SLOPe <Type>
This command selects the slope of the trigger signal. The selected trigger slope
applies to all trigger signal sources.
Suffix:
<n>
.
irrelevant
Parameters:
<Type> POSitive | NEGative
*RST: POSitive
Example:
TRIG:SLOP NEG
Manual operation: See "Trg/Gate Polarity" on page 272
See "Trigger Polarity" on page 462
TRIGger<n>[:SEQuence]:SOURce <Source>
This command selects the trigger source.
Note: Using gating with gate mode "level" ( TRACe<n>:IQ:EGATe:TYPE
on page 899/[SENSe:]SWEep:EGATe:TYPE) and an IFP trigger, the holdoff time for
the IFP trigger (see TRIGger<n>[:SEQuence]:HOLDoff[:TIME] on page 906) is
ignored for frequency sweep, FFT sweep, zero span and IQ mode measurements.
When triggering or gating is activated, the squelch function is automatically disabled
(see [SENSe:]DEMod:SQUelch[:STATe]).
IF power and RF power triggers are not available together with the bandwidth exten-
sion option R&S FSV-B160.
For details on trigger modes refer to "Trg/Gate Source" on page 268.
For details on trigger modes refer to the "Trg/Gate Source" softkey in the base unit
description.
Suffix:
<n>
.
irrelevant
Remote Control – Commands