Remote Control
R&S
®
FSVA/FSV
910Operating Manual 1176.7510.02 ─ 09
Parameters:
<Source> IMMediate
Free Run
EXTern
External trigger
IFPower
Power trigger at the second intermediate frequency
RFPower
Power trigger at the first intermediate frequency
GP0 | GP1 | GP2 | GP3 | GP4 | GP5
For I/Q Analyzer or AnalogDemod mode only:
Defines triggering of the measurement directly via the LVDS
connector. The parameter specifies which general purpose bit (0
to 5) will provide the trigger data.
This trigger mode is available for input from the R&S Digital I/Q
Interface (option R&S FSV-B17) only.
The assignment of the general purpose bits used by the Digital
IQ trigger to the LVDS connector pins is provided in Table 4-19
(See also Table 3-21).
VIDeo
Video mode is only available in the time domain and only in
Spectrum mode.
BBPower
Baseband power (for digital input via the R&S Digital I/Q Inter-
face, R&S FSV-B17)
PSEN
External power sensor (requires R&S FSV-K9 option)
*RST:
IMMediate
Example:
TRIG:SOUR EXT
Selects the external trigger input as source of the trigger signal
Manual operation: See "Trg/Gate Source" on page 268
See "Free Run" on page 269
See "External" on page 269
See "Video" on page 269
See "RF Power" on page 269
See "IF Power/BB Power" on page 270
See "Power Sensor" on page 270
See "Time" on page 271
See "Digital IQ" on page 271
Table 4-19: Assignment of general purpose bits to LVDS connector pins
Bit LVDS pin
GP0 SDATA4_P - Trigger1
GP1 SDATA4_P - Trigger2
GP2 SDATA0_P - Reserve1
Remote Control – Commands