Instrument Functions
R&S
®
FSVA/FSV
272Operating Manual 1176.7510.02 ─ 09
Bit LVDS pin
GP2 SDATA0_P - Reserve1
GP3 SDATA4_P - Reserve2
GP4 SDATA0_P - Marker1
GP5 SDATA4_P - Marker2
Remote command:
TRIG:SOUR GP0, see TRIGger<n>[:SEQuence]:SOURce on page 909
SWE:EGAT:SOUR RFP for gated triggering, see [SENSe:]SWEep:EGATe:SOURce
on page 854
Trg/Gate Level
Opens an edit dialog box to enter the trigger/gate level.
For details see also Chapter 3.2.9.3, "Using Gated Sweep Operation", on page 276.
In the trigger modes "Time" and "Power Sensor", this softkey is not available.
Remote command:
TRIGger<n>[:SEQuence]:LEVel:IFPower on page 908
TRIGger<n>[:SEQuence]:LEVel:VIDeo on page 908
Trg/Gate Polarity
Sets the polarity of the trigger/gate source.
The sweep starts after a positive or negative edge of the trigger signal. The default set-
ting is "Pos". The setting applies to all trigger modes with the exception of the "Free
Run", "Power Sensor" and "Time" mode.
For details also see Chapter 3.2.9.3, "Using Gated Sweep Operation", on page 276.
"Pos"
Level triggering: the sweep is stopped by the logic "0" signal and
restarted by the logical "1" signal after the gate delay time has
elapsed.
"Neg"
Edge triggering: the sweep is continued on a "0" to "1" transition for
the gate length duration after the gate delay time has elapsed.
Remote command:
TRIGger<n>[:SEQuence]:SLOPe on page 909
[SENSe:]SWEep:EGATe:POLarity on page 854
Trigger Offset
Opens an edit dialog box to enter the time offset between the trigger signal and the
start of the sweep.
Measurement Parameters