Remote Control
R&S
®
FSVA/FSV
906Operating Manual 1176.7510.02 ─ 09
TRIGger<n>[:SEQuence]:SLOPe.....................................................................................909
TRIGger<n>[:SEQuence]:SOURce.................................................................................. 909
TRIGger<n>[:SEQuence]:TIME:RINTerval........................................................................911
TRIGger<n>[:SEQuence]:HOLDoff[:TIME] <Delay>
This command defines the length of the trigger delay.
A negative delay time (pretrigger) can be set in zero span only.
Suffix:
<n>
.
irrelevant
Parameters:
<Delay> Range: zero span: -sweeptime (see data sheet) to 30 s;
span: 0 to 30 s
*RST: 0 s
Example:
TRIG:HOLD 500us
Manual operation: See "Trigger Offset" on page 272
TRIGger<n>[:SEQuence]:BBPower:HOLDoff <Value>
This command sets the holding time before the next BB power trigger event (for digital
input via the R&S Digital I/Q Interface, R&S FSV-B17).
Suffix:
<n>
.
irrelevant
Parameters:
<Value> *RST: 150 ns
Example:
TRIG:SOUR BBP
Sets the baseband power trigger source.
TRIG:BBP:HOLD 200 ns
Sets the holding time to 200 ns.
Mode: all
Manual operation: See "Trigger Holdoff" on page 273
TRIGger<n>[:SEQuence]:IFPower:HOLDoff <Value>
This command sets the holding time before the next IF power trigger event.
Note: Using gating with gate mode "level" (see TRACe<n>:IQ:EGATe:TYPE
on page 899) and an IFP trigger (see TRIGger<n>[:SEQuence]:LEVel:IFPower
on page 908), the holdoff time for the IFP trigger is ignored for frequency sweep, FFT
sweep, zero span and IQ mode measurements.
Suffix:
<n>
.
irrelevant
Remote Control – Commands