Remote Control
R&S
®
FSVA/FSV
907Operating Manual 1176.7510.02 ─ 09
Parameters:
<Value> *RST: 150 ns
Example:
TRIG:SOUR IFP
Sets the IF power trigger source.
TRIG:IFP:HOLD 200 ns
Sets the holding time to 200 ns.
Manual operation: See "Trigger Holdoff" on page 273
TRIGger<n>[:SEQuence]:IFPower:HYSTeresis <Value>
This command sets the limit that the hysteresis value for the IF power trigger has to fall
below in order to trigger the next measurement.
Suffix:
<n>
.
irrelevant
Parameters:
<Value> *RST: 3 dB
Example:
TRIG:SOUR IFP
Sets the IF power trigger source.
TRIG:IFP:HYST 10DB
Sets the hysteresis limit value.
Manual operation: See "Trigger Hysteresis" on page 273
TRIGger<n>[:SEQuence]:LEVel:BBPower <Level>
This command sets the level of the baseband power trigger source (for digital input via
the R&S Digital I/Q Interface, R&S FSV-B17).
Suffix:
<n>
.
irrelevant
Parameters:
<Level> Range: -50 dBm to +20 dBm
*RST: -20 DBM
Example:
TRIG:LEV:BB -30DBM
Mode: All
Manual operation: See "Trigger Level" on page 462
TRIGger<n>[:SEQuence]:LEVel[:EXTernal] <TriggerLevel>
This command sets the level of the external trigger source in Volt.
Suffix:
<n>
.
irrelevant
Remote Control – Commands