Remote Control
R&S
®
FSVA/FSV
899Operating Manual 1176.7510.02 ─ 09
Parameters:
<GateLength> <numeric value>
Max = MaxPreTriggerSamples * 128 MHz/sample rate
pretrigger samples defined by TRACe<n>:IQ:SET;
sample rate defined by TRACe<n>:IQ:SRATe)
Range: 1...Max (samples)
*RST: 100
Example:
TRAC:IQ:EGAT:LENG 2000
TRACe<n>:IQ:EGATe:NOFgateperiods <Number>
Defines the number of gate periods after a trigger signal. For details see "I/Q Gating"
on page 904.
This function is not available if the R&S Digital I/Q Interface (R&S FSV-B17) is active.
Suffix:
<n>
.
irrelevant
Parameters:
<Number> 1...1023
*RST: 1
Example:
TRAC:IQ:EGAT:NOF 2
TRACe<n>:IQ:EGATe:TYPE <Type>
Specifies the gate trigger type used for data collection. For details see "I/Q Gating"
on page 904.
Note: Using gating with gate mode "level" and an IFP trigger ( TRIGger<n>[:
SEQuence]:SOURce), the holdoff time for the IFP trigger (see TRIGger<n>[:
SEQuence]:IFPower:HOLDoff on page 906) is ignored.
This function is not available if the R&S Digital I/Q Interface (R&S FSV-B17) is active.
Suffix:
<n>
.
irrelevant
Parameters:
<Type> LEVel | EDGE
*RST: EDGE
Example:
TRAC:IQ:EGAT:TYPE LEV
TRACe<n>:IQ:EVAL <State>
This command enables or disables the IQ data evaluation mode and is a prerequisite
for the IQ Analyzer functions.
Remote Control – Commands