Remote Control
R&S
®
FSVA/FSV
898Operating Manual 1176.7510.02 ─ 09
Mode: IQ
Manual operation: See "No Filter" on page 446
TRACe<n>:IQ:EGATe <State>
Activates or deactivates the gate function for IQ data measurements. This command is
only valid for TRAC:IQ ON and trigger modes "EXT" and "IFP". The gate trigger is
specified using TRACe<n>:IQ:SET on page 901. For details see "I/Q Gating"
on page 904.
This function is not available if the R&S Digital I/Q Interface (R&S FSV-B17) is active.
Suffix:
<n>
.
irrelevant
Parameters:
<State> ON | OFF
*RST: OFF
Example:
TRAC:IQ:EGAT: ON
TRACe<n>:IQ:EGATe:GAP <Samples>
Defines the interval between several gate periods in samples. For details see "I/Q Gat-
ing" on page 904.
This function is not available if the R&S Digital I/Q Interface (R&S FSV-B17) is active.
Suffix:
<n>
.
irrelevant
Parameters:
<Samples> <numeric value>
Max = MaxPreTriggerSamples * 128 MHz/sample rate
pretrigger samples defined by TRACe<n>:IQ:SET;
sample rate defined by TRACe<n>:IQ:SRATe)
Range: 1...Max (samples)
*RST: 100
Example:
TRAC:IQ:EGAT:GAP 2
TRACe<n>:IQ:EGATe:LENGth <GateLength>
Defines the gate length in samples in edge mode. For details see "I/Q Gating"
on page 904.
This function is not available if the R&S Digital I/Q Interface (R&S FSV-B17) is active.
Suffix:
<n>
.
irrelevant
Remote Control – Commands