Remote Control
R&S
®
FSVA/FSV
866Operating Manual 1176.7510.02 ─ 09
This command is requires the optional External Tracking Generator.
Suffix:
<n>
.
irrelevant
<generator> 1 | 2
external generator
Parameters:
<State> ON | OFF
*RST: OFF
Example:
SOUR:EXT1:FREQ:SWE ON
Activates the frequency sweep for external generator 1.
SOURce<n>:EXTernal<generator>:POWer[:LEVel] <Level>
This command sets the output power of the selected generator.
This command is only valid for R&S FSVA/FSV option External Tracking Generator
(R&S FSV-B10).
Suffix:
<n>
.
irrelevant
<generator> 1 | 2
tracking generator
For Noise Figure measurements (K30): 1 only
Parameters:
<Level> <numeric value>
*RST: -20 dBm
Example:
SOUR:EXT:POW -30dBm
Sets the generator level to -30 dBm
Mode: A, NF
Manual operation: See "Source Power" on page 487
SOURce<n>:EXTernal<generator>:ROSCillator[:SOURce] <Source>
This command controls selection of the reference oscillator for the external tracking
generator (requires option R&S FSV-B10).
If the external reference oscillator is selected, the reference signal must be connected
to the rear panel of the instrument.
Suffix:
<n>
.
irrelevant
<generator> 1 | 2
tracking generator
For Noise Figure measurements (K30): 1 only
Remote Control – Commands