Remote Control
R&S
®
FSVA/FSV
867Operating Manual 1176.7510.02 ─ 09
Parameters:
<Source> INTernal
the internal reference is used
EXTernal
the external reference is used; if none is available, an error flag
is displayed in the status bar
EAUTo
the external reference is used as long as it is available, then the
instrument switches to the internal reference (Spectrum mode
only)
*RST: INT
Example:
SOUR:EXT:ROSC EXT
Switches to external reference oscillator
Mode: A, NF
Manual operation: See "Reference" on page 489
SOURce<n>:FM:DEViation <Frequency>
This command defines the maximum frequency deviation at 1 V input voltage at the
FM input of the tracking generator.
This command is available with a tracking generator (option R&S FSV-B9).
Suffix:
<n>
.
1...4
Selects the measurement window.
Parameters:
<Frequency> The possible value range is 100 Hz to 10 MHz in steps of 10 Hz.
Range: 100 Hz to 10 MHz
Increment: 10 Hz
*RST: 100 Hz
Example:
SOUR1:FM:DEV 1 MHz
Sets the maximum frequency deviation of the tracking generator
for screen A to 1 MHz.
Usage: SCPI confirmed
Manual operation: See "External FM" on page 478
SOURce<n>:FM:STATe <State>
This command turns external frequency modulation of the tracking generator on and
off.
External I/Q modulation is switched off, if active. This command is available with a
tracking generator (option R&S FSV-B9).
Remote Control – Commands