Remote Control
R&S
®
FSVA/FSV
862Operating Manual 1176.7510.02 ─ 09
SOURce<n>:DM:STATe <State>
This command turns external I/Q modulation of the tracking generator on and off.
External AM and external FM are switched off, if active. This command is available
with a tracking generator (option R&S FSV-B9).
Suffix:
<n>
.
1...4
Selects the measurement window.
Parameters:
<State> ON | OFF
*RST: OFF
Example: SOUR2:DM:STAT ON
Switches on the external I/Q modulation of the tracking genera-
tor for screen B.
Usage: SCPI confirmed
Manual operation: See "External I/Q" on page 478
See "Modulation OFF" on page 478
See "External IQ" on page 497
SOURce<n>:EXTernal<generator>[:STATe] <State>
This command activates or deactivates the external generator indicated by the suffix
<generator>.
This command is requires the optional External Tracking Generator.
Suffix:
<n>
.
irrelevant
<generator> 1 | 2
external generator
Parameters:
<State> ON | OFF
*RST: OFF
Example:
SYST:COMM:RDEV:GEN1:TYPE 'SMP02'
Selects SMP02 as generator 1.
"SYST:COMM:RDEV:GEN1:LINK TTL"
Selects GPIB + TTL link as interface.
"SYST:COMM:RDEV:GEN1:ADDR 28"
Sets the generator address to 28.
"SOUR:EXT1:FREQ:SWE ON"
Activates the frequency sweep for generator 1.
"SOUR:EXT ON"
Activates the external generator
Manual operation: See "Source RF Internal (On/ Off)" on page 475
Remote Control – Commands