Remote Control
R&S
®
FSVA/FSV
863Operating Manual 1176.7510.02 ─ 09
SOURce<n>:EXTernal<generator>:FREQuency <Frequency>
This command defines a fixed source frequency for the external tracking generator.
This command is requires the optional External Tracking Generator.
Suffix:
<n>
.
1...4
window
<m> 1 | 2
tracking generator
Parameters:
<Frequency> Source frequency of the external tracking generator.
*RST: RST value
Example:
SOUR:EXT2:FREQ 10MHz
Manual operation: See "(Fixed) Generator Frequency" on page 486
SOURce<n>:EXTernal<generator>:FREQuency:COUPling[:STATe]
<State>
This command couples or decouples the analyzer and generator frequencies.
This command is requires the optional External Tracking Generator.
Suffix:
<generator>
.
1 | 2
tracking generator
Parameters:
<State> ON | OFF
ON
OFF
The fixed frequency defined using SOURce<n>:
EXTernal<generator>:FREQuency is used as a source fre-
quency.
*RST: OFF
Example:
SOUR:EXT2:FREQ:COUP ON
Manual operation: See "Source Frequency Coupling" on page 486
SOURce<n>:EXTernal<generator>:FREQuency[:FACTor]:DENominator <Value>
This command defines the denominator of the factor with which the analyzer frequency
is multiplied in order to obtain the transmit frequency of the selected generator.
This command is only valid for R&S FSVA/FSV option External Tracking Generator
(R&S FSV-B10).
Remote Control – Commands