Instrument Functions
R&S
®
FSVA/FSV
255Operating Manual 1176.7510.02 ─ 09
If a detector was selected manually, the "MAN" indicator is highlighted.
If "AUTO" is selected, the detector is defined automatically, depending on the selected
trace mode:
Trace mode Detector
Clear Write Auto Peak
Max Hold Positive Peak
Min Hold Negative Peak
Average Sample Peak
View
Blank
For details see Chapter 3.2.8.6, "Detector Overview", on page 265.
Auto Select ← Detector ← Trace 1/Trace 2/Trace 3/Trace 4/Trace 5/Trace 6
Selects the best detector for the selected trace and filter mode. This is the default set-
ting.
For details see also Chapter 3.2.8.6, "Detector Overview", on page 265.
Trace mode Detector
Clear/Write Auto Peak
Average Sample
Max Hold Max Peak
Min Hold Min Peak
Remote command:
[SENSe:][WINDow:]DETector<trace>[:FUNCtion]:AUTO on page 860
Auto Peak ← Detector ← Trace 1/Trace 2/Trace 3/Trace 4/Trace 5/Trace 6
Selects the "Auto Peak" detector.
For details see Chapter 3.2.8.6, "Detector Overview", on page 265.
Remote command:
DET APE, see [SENSe:][WINDow:]DETector<trace>[:FUNCtion]
on page 860
Positive Peak ← Detector ← Trace 1/Trace 2/Trace 3/Trace 4/Trace 5/Trace 6
Selects the "Positive Peak" detector.
For details see Chapter 3.2.8.6, "Detector Overview", on page 265.
Remote command:
DET POS, see [SENSe:][WINDow:]DETector<trace>[:FUNCtion]
on page 860