Instrument Functions
R&S
®
FSVA/FSV
255Operating Manual 1176.7510.02 ─ 09
If a detector was selected manually, the "MAN" indicator is highlighted.
If "AUTO" is selected, the detector is defined automatically, depending on the selected
trace mode:
Trace mode Detector
Clear Write Auto Peak
Max Hold Positive Peak
Min Hold Negative Peak
Average Sample Peak
View –
Blank –
For details see Chapter 3.2.8.6, "Detector Overview", on page 265.
Auto Select ← Detector ← Trace 1/Trace 2/Trace 3/Trace 4/Trace 5/Trace 6
Selects the best detector for the selected trace and filter mode. This is the default set-
ting.
For details see also Chapter 3.2.8.6, "Detector Overview", on page 265.
Trace mode Detector
Clear/Write Auto Peak
Average Sample
Max Hold Max Peak
Min Hold Min Peak
Remote command:
[SENSe:][WINDow:]DETector<trace>[:FUNCtion]:AUTO on page 860
Auto Peak ← Detector ← Trace 1/Trace 2/Trace 3/Trace 4/Trace 5/Trace 6
Selects the "Auto Peak" detector.
For details see Chapter 3.2.8.6, "Detector Overview", on page 265.
Remote command:
DET APE, see [SENSe:][WINDow:]DETector<trace>[:FUNCtion]
on page 860
Positive Peak ← Detector ← Trace 1/Trace 2/Trace 3/Trace 4/Trace 5/Trace 6
Selects the "Positive Peak" detector.
For details see Chapter 3.2.8.6, "Detector Overview", on page 265.
Remote command:
DET POS, see [SENSe:][WINDow:]DETector<trace>[:FUNCtion]
on page 860
Measurement Parameters