Remote Control
R&S
®
FSVA/FSV
837Operating Manual 1176.7510.02 ─ 09
Parameters:
<Duration> SHORt | NORMal | LONG
*RST: NORM
Example:
PMET2:MTIM SHOR
Sets a short measurement duration for measurements of station-
ary high power signals for the selected power sensor.
Mode: PSM
Manual operation: See "Meas Time / Average" on page 548
[SENSe:]PMETer<p>:MTIMe:AVERage[:STATe] <State>
This command activates or deactivates manual averaging for the selected power sen-
sor. The average count is set with the [SENSe:]PMETer<p>:MTIMe:AVERage:
COUNt command.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<State> ON | OFF
*RST: OFF
Example:
PMET2:MTIM:AVER ON
Activates manual averaging.
Mode: PSM
Manual operation: See "Meas Time / Average" on page 548
[SENSe:]PMETer<p>:MTIMe:AVERage:COUNt <NumberReadings>
This command defines the number of readings (averagings) to be performed after a
single sweep has been started for the selected power sensor. This command is only
available if the [SENSe:]PMETer<p>:MTIMe:AVERage[:STATe] command is set to
ON.
Results become more stable with extended average, particularly if signals with low
power are measured. This setting can be used to minimize the influence of noise in the
power meter measurement.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<NumberReadings> 0 to 256 in binary steps (1, 2, 4, 8,…)
For average count = 0 or 1, one reading is performed.
Remote Control – Commands