Remote Control
R&S
®
FSVA/FSV
838Operating Manual 1176.7510.02 ─ 09
Example:
PMET2:MTIM:AVER ON
Activates manual averaging.
PMET2:MTIM:AVER:COUN 8
Sets the number of readings to 8.
Mode: PSM
Manual operation: See "Number of Readings" on page 549
[SENSe:]PMETer<p>:ROFFset[:STATe] <State>
This command defines whether the reference level offset set for the analyzer is taken
into account for the measured power or not for the selected power sensor.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<State> ON
Adds the offset defined by the DISPlay[:WINDow<n>]:
TRACe<t>:Y[:SCALe]:RLEVel:OFFSet command to the
measured level.
OFF
Takes no offset into account.
*RST: ON
Example:
PMET2:ROFF OFF
Takes no offset into account for the measured power.
Mode: PSM
Manual operation: See "Use Ref Level Offset" on page 549
[SENSe:]PMETer<p>:TRIGger[:STATe] <State>
This command switches the external power trigger on or off for the selected power
sensor.
For details on configuring an external power trigger and supported power sensors, see
Chapter 3.9.3, "Configuring an External Power Trigger", on page 543.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<State> ON | OFF
*RST: OFF
Example:
PMET2:TRIG ON
Switches the external power trigger on
Mode: PSM
Remote Control – Commands