Remote Control
R&S
®
FSVA/FSV
751Operating Manual 1176.7510.02 ─ 09
Parameters:
<ReferenceLevel> The unit is variable.
Range: see datasheet
*RST: -10dBm
Example:
DISP:TRAC:Y:RLEV -60dBm
Manual operation: See "Ref Level" on page 226
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel:OFFSet <Value>
This command defines a reference level offset.
Suffix:
<n>
.
irrelevant.
<t> irrelevant
Parameters:
<Value> Range: -200 to 200
*RST: 0
Default unit: dB
Example:
DISP:TRAC:Y:RLEV:OFFS -10dB
Manual operation: See "Ref Level Offset" on page 230
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RPOSition <Position>
This command defines the position of the reference level on the display grid..
When using a tracking generator (only with option R&S FSV-B9 or -B10, requires
active normalization), and in Bluetooth mode (option R&S FSV-K8) this command
defines the position of the reference value for all windows.
Suffix:
<n>
.
Selects the measurement window.
<t> irrelevant
Parameters:
<Position> 0 PCT corresponds to the lower display border, 100% corre-
sponds to the upper display border.
Range: 0 to 100
*RST: Spectrum mode: 100 PCT, with tracking generator
or time display: 50 PCT
Default unit: PCT
Example:
DISP:TRAC:Y:RPOS 50PCT
Manual operation: See "Ref Level Position" on page 230
See "Reference Value Position" on page 477