Remote Control
R&S
®
FSVA/FSV
752Operating Manual 1176.7510.02 ─ 09
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RVALue <Value>
The command defines the power value assigned to the reference position in the grid.
When using a tracking generator, this command requires active normalization.
Suffix:
<n>
.
irrelevant
<t> irrelevant
Parameters:
<Value> *RST: 0 dB, coupled to reference level
Example:
DISP:TRAC:Y:RVAL -20dBm
Defines a reference position of -20 dBm.
Manual operation: See "Reference Level" on page 445
See "Reference Value" on page 477
DISPlay[:WINDow<n>]:ZOOM:AREA
<x1>, <y1>, <x2>, <y2>
This command defines the zoom area.
Before you can define a zoom area, you first have to turn the zoom on.
Suffix:
<n>
.
window; For applications that do not have more than 1 measure-
ment window, the suffix <n> is irrelevant.
Parameters:
<x1>,<y1>,
<x2>,<y2>
Diagram coordinates in % of the complete diagram that define
the zoom area.
The lower left corner is the origin of coordinate system. The
upper right corner is the end point of the system.
Range: 0 to 100
Default unit: PCT
Example:
DISP:ZOOM ON
Activates the zoom mode.
DISP:ZOOM:AREA 5,30,20,100
Enlarges the display of the measurement results in the area
defined by the coordinates (5,30) and (20,100).
Remote Control – Commands