Instrument Functions
R&S
®
FSVA/FSV
445Operating Manual 1176.7510.02 ─ 09
If you select the "File > Exit" menu item in the R&S DigIConf window, the application is
closed. Note that in this case the settings are lost and the EX-IQ-BOX functionality is
no longer available until you restart the application using the "DigIConf" softkey in the
R&S FSVA/FSV once again.
Remote command:
Remote commands for the R&S DigIConf software always begin with SOURce:EBOX.
Such commands are passed on from the R&S FSVA/FSV to the R&S DigIConf auto-
matically which then configures the R&S EX-IQ-BOX via the USB connection.
All remote commands available for configuration via the R&S DigIConf software are
described in the "R&S®EX-IQ-BOX Digital Interface Module R&S®DigIConf Software
Operating Manual".
Example 1:
SOURce:EBOX:*RST
SOURce:EBOX:*IDN?
Result:
"Rohde&Schwarz,DigIConf,02.05.436 Build 47"
Example 2:
SOURce:EBOX:USER:CLOCk:REFerence:FREQuency 5MHZ
Defines the frequency value of the reference clock.
Level ← I/Q Analyzer
Opens a dialog box to define the level settings.
Reference Level ← Level ← I/Q Analyzer
Specifies the reference level for the I/Q measurement.
Remote command:
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RVALue on page 752
Auto Level ← Level ← I/Q Analyzer
If enabled, the reference level is adjusted to the full scale level automatically if the full
scale level changes.
Remote command:
INPut:DIQ:RANGe:COUPling on page 761
Preamp On/Off ← Level ← I/Q Analyzer
Switches the preamplifier on and off.
If option R&S FSV-B22 is installed, the preamplifier is only active below 7 GHz.
If option R&S FSV-B24 is installed, the preamplifier is active for all frequencies.
When measuring Spurious Emissions, using this softkey automatically opens the
"Sweep List" dialog, see "Sweep List dialog box" on page 378.
This function is not available for input from the R&S Digital I/Q Interface (option
R&S FSV-B17).
Remote command:
INPut:GAIN:STATe on page 764
Data Acquisition ← I/Q Analyzer
Opens a dialog box to configure data acquisition in I/Q Analyzer mode.
Instrument Functions - I/Q Analyzer