Remote Control
R&S
®
FSVA/FSV
761Operating Manual 1176.7510.02 ─ 09
<PRBSTestState> State of the PRBS test.
Not Started
Has to be Started
Started
Passed
Failed
Done
<SampleRateType> 0
Maximum sampling rate is displayed
1
Current sampling rate is displayed
<Placeholder> for future use; currently "0"
Example:
INP:DIQ:CDEV?
Result:
1,SMU200A,103634,Out
A,70000000,100000000,Passed,Not Started,0,0
Mode: IQ, VSA, EVDO, CDMA, WCDMA, GSM, ADEMOD, TDS
Manual operation: See "Connected Device" on page 429
See "Digital IQ Info" on page 432
INPut:DIQ:RANGe:COUPling <State>
If enabled, the reference level for digital input is adjusted to the full scale level automat-
ically if the fullscale level changes.
(See the Auto Level softkey).
This command is only available if the optional R&S Digital I/Q Interface (option
R&S FSV-B17) is installed.
For details see the R&S Digital I/Q Interface (R&S FSV-B17) description of the base
unit.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
INP:DIQ:RANG:COUP OFF
Mode: IQ, VSA, EVDO, CDMA, WCDMA, GSM, ADEMOD, TDS
Manual operation: See "Adjust Reference Level to Full Scale Level" on page 430
See "Auto Level" on page 445
INPut:DIQ:RANGe[:UPPer] <Level>
Defines or queries the "Full Scale Level", i.e. the level that should correspond to an I/Q
sample with the magnitude "1".
Remote Control – Commands