Remote Control
R&S
®
FSVA/FSV
762Operating Manual 1176.7510.02 ─ 09
It can be defined either in dBm or Volt (see "Full Scale Level" on page 430).
This command is only available if the optional R&S Digital I/Q Interface (option
R&S FSV-B17) is installed.
For details see Chapter 3.7.5, "Interface Status Information", on page 513.
Parameters:
<Level> <numeric value>
Range: 70.711 nV to 7.071 V
*RST: 1 V
Example:
INP:DIQ:RANG 1V
Mode: A, IQ, NF, TDS, VSA, CDMA, EVDO, WCDMA, ADEMOD,
GSM, OFDM, OFDMA/WiBro, WLAN
Manual operation: See "Full Scale Level" on page 430
INPut:DIQ:RANGe[:UPPer]:UNIT <Unit>
Defines the unit of the full scale level (see "Level Unit" on page 430). The availability of
units depends on the measurement application you are using.
This command is only available if the optional R&S Digital I/Q Interface (option
R&S FSV-B17) is installed.
For details see the R&S Digital I/Q Interface (R&S FSV-B17) description of the base
unit.
Parameters:
<Level> V | dBm | dBpW | W | dBmV | dBuV | dBuA | A
*RST: Volt
Example:
INP:DIQ:RANG:UNIT A
Mode: IQ, VSA, EVDO, CDMA, WCDMA, GSM, ADEMOD, TDS
Manual operation: See "Level Unit" on page 430
INPut:DIQ:SRATe <SampleRate>
This command specifies or queries the sample rate of the input signal from the R&S
Digital I/Q Interface (see "Input Sample Rate" on page 429).
Note: the final user sample rate of the R&S FSVA/FSV may differ and is defined using
TRAC:IQ:SRAT (see TRACe<n>:IQ:SRATe on page 903).
This command is only available if the optional R&S Digital I/Q Interface (option
R&S FSV-B17) is installed.
For details see the R&S Digital I/Q Interface (R&S FSV-B17) description of the base
unit.
Remote Control – Commands