Remote Control
R&S
®
FSVA/FSV
750Operating Manual 1176.7510.02 ─ 09
<t> irrelevant
Parameters:
<Range> Range: 10 to 200
*RST: 100
Default unit: dB
Example:
DISP:TRAC:Y 110dB
Manual operation: See "Range Log 100 dB" on page 226
See "Range Log 50 dB" on page 226
See "Range Log 10 dB" on page 227
See "Range Log 5 dB" on page 227
See "Range Log 1 dB" on page 227
See "Range Log Manual" on page 227
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:MODE
<Mode>
This command selects the type of scaling of the y-axis.
When SYSTem:DISPlay:UPDate is turned off, this command has no immediate
effect on the screen.
Suffix:
<n>
.
window; For applications that do not have more than 1 measure-
ment window, the suffix <n> is irrelevant.
<t> irrelevant
Parameters:
<Mode> ABSolute
absolute scaling of the y-axis
RELative
relative scaling of the y-axis
*RST: ABS
Example:
DISP:TRAC:Y:MODE REL
Manual operation: See "Grid Abs/Rel " on page 230
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel <ReferenceLevel>
This command defines the reference level.
With the reference level offset 0, the value range of the reference level is modified by
the offset.
The unit depends on the setting defined with CALCulate<n>:UNIT:POWer.
Suffix:
<n>
.
irrelevant.
<t> irrelevant
Remote Control – Commands