Remote Control
R&S
®
FSVA/FSV
749Operating Manual 1176.7510.02 ─ 09
Parameters:
<ScalingType> LOGarithmic | LINear
LOGarithmic
Selects logarithmic scaling.
LINear
Selects linear scaling.
*RST: LINear
Example:
DISP:TRAC:X:SPAC LOG
Mode: A, ADEMOD
DISPlay[:WINDow<n>]:TRACe<t>:Y:SPACing
<ScalingType>
This command selects the scaling of the y-axis.
Suffix:
<n>
.
Selects the measurement window.
<t> irrelevant
Parameters:
<ScalingType> LOGarithmic
Logarithmic scaling.
LINear
Linear scaling in %.
LDB
Linear scaling in dB.
*RST: LOGarithmic
Example:
DISP:TRAC:Y:SPAC LIN
Select a linear scale.
Manual operation: See "Range Log 100 dB" on page 226
See "Range Log 50 dB" on page 226
See "Range Log 10 dB" on page 227
See "Range Log 5 dB" on page 227
See "Range Log 1 dB" on page 227
See "Range Log Manual" on page 227
See "Range Linear %" on page 227
See "Range Lin. Unit" on page 228
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe] <Range>
This command defines the display range of the y-axis with logarithmic scaling.
The command works only for a logarithmic scaling. You can select the scaling with
DISPlay[:WINDow<n>]:TRACe<t>:Y:SPACing on page 749.
Suffix:
<n>
.
window; For applications that do not have more than 1 measure-
ment window, the suffix <n> is irrelevant.
Remote Control – Commands

Table of Contents