Remote Control
R&S
®
FSVA/FSV
839Operating Manual 1176.7510.02 ─ 09
Manual operation: See "External Power Trigger" on page 549
[SENSe:]PMETer<p>:TRIGger:DTIMe <Value>
Defines the time the input signal must stay below the IF power trigger level in order to
allow a trigger to start the measurement.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<Value> 0 to 1 s
Increment: 100 ns
*RST: 100 µs
Example:
PMET2:TRIG:DTIMe 0.001
Mode: PSM
Manual operation: See "Dropout" on page 550
[SENSe:]PMETer<p>:TRIGger:HOLDoff <Value>
Defines the holdoff value in s, which is the time which must pass before triggering, in
case another trigger event happens.
For details see Figure 3-47.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<Value> 0 to 1 s
Increment: 100 ns
*RST: 0 s
Example:
PMET2:TRIG:HOLD 0.1
Sets the holdoff time of the trigger to 100 ms
Mode: PSM
Manual operation: See "Holdoff" on page 550
[SENSe:]PMETer<p>:TRIGger:HYSTeresis <Value>
Defines the value for the trigger hysteresis of an external power trigger. The hysteresis
in dB is the value the input signal must stay below the IF power trigger level in order to
allow a trigger to start the measurement.
For details see Figure 3-47.