836Operating Manual 1176.7510.02 ─ 09
Manual operation: See "Duty Cycle" on page 548
[SENSe:]PMETer<p>:FREQuency <Frequency>
This command sets the frequency of the selected power sensor. The limits f
min
and f
max
are defined by the power sensor.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<Frequency> fmin ... fmax
*RST: 50 MHz
Example:
PMET2:FREQ 1GHZ
Sets the frequency of the power sensor to 1 GHz.
Mode: PSM
Manual operation: See "Frequency Manual" on page 547
[SENSe:]PMETer<p>:FREQuency:LINK <Coupling>
This command sets the frequency coupling of the selected power sensor.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<Coupling> CENTer
Couples the frequency to the center frequency of the analyzer
MARKer1
couples the frequency to the position of marker 1
OFF
switches the frequency coupling off
*RST: CENT
Example:
PMET2:FREQ:LINK CENT
Couples the frequency to the center frequency of the analyzer
Mode: PSM
Manual operation: See "Frequency Coupling" on page 548
[SENSe:]PMETer<p>:MTIMe <Duration>
This command sets the measurement duration of the selected power sensor.
Suffix:
<p>
.
1...4
Power sensor index
Remote Control – Commands