Remote Control
R&S
®
FSVA/FSV
835Operating Manual 1176.7510.02 ─ 09
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<State> ON | OFF
*RST: OFF
Example:
PMET1 ON
Switches the power sensor measurements on.
Mode: PSM
Manual operation: See "Power Sensor On/Off" on page 545
See "Select" on page 547
[SENSe:]PMETer<p>:DCYCle:STATe
<State>
This command activates the duty cycle correction for the selected power sensor.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<State> ON | OFF
*RST: OFF
Example:
PMET2:DCYC:STAT ON
Mode: PSM
Manual operation: See "Duty Cycle" on page 548
[SENSe:]PMETer<p>:DCYCle:VALue <Percentage>
This command sets the duty cycle for the selected power sensor to a percent value for
the correction of pulse modulated signals. With the correction activated, the sensor cal-
culates the signal pulse power from this value and the mean power.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<Percentage> Range: 0.001 to 99.999
*RST: 99.999
Default unit: %
Example:
PMET2:DCYC:STAT ON
Activates the duty cycle correction.
PMET2:DCYC:VAL 0.5
Sets the correction value to 0.5%.
Mode: PSM
Remote Control – Commands